Changing The Face Of Blood-Pressure Monitoring



Roger Allan  |   ED Online ID #11520  |   December 1, 2005

Article Rating: Not Rated

The word "tensymetry" has no definition in Webster's Dictionary, but it's well known in the medical world. Developed by Tensys Medical Systems, tensymetry is a technology that uses a proprietary combination of biomechanical, electrical, and software engineering. With that powerful trio, it can perform accurate, continuous, real-time, and non-invasive measurement of a patient's beat-to-beat blood pressure in an operating room.

The fruit borne from that technology has been the company's T-line Tensymeters. The most recent of this line, the TL-150, was introduced this March (Fig. 1). Worn on a patient's cuff, the TL-150 vastly improves on its predecessors thanks to advances in motion tolerance. It measures blood pressure at the radial artery.

According to leading anesthesiologists, this development is nothing short of revolutionary. "The TL-150 will change the way we measure blood pressure in the future," says Dr. Peter Szmuk, associate professor of anesthesiology at the University of Texas Medical School, Houston. "Results show remarkable correlation between the T-Line and ALine (arterial line) reference during clinical situations in which a typical non-invasive blood-pressure (NIBP) cuff is known to be highly suspect and beat-to-beat vigilance is highly desirable."

The A-line reference is an FDA-approved standard (SP10:2002, accuracy specification 4.4.5.2.B) established by the Association for Advanced Medical Instrumentation (AAMI) for non-invasive blood-pressure monitors. It specifies a mean difference standard deviation within ±5 mm Hg (mercury), not to exceed 8 mm Hg, which the TL-150 meets. A case study confirms the T-Line's importance versus the use of infrequent non-invasive blood-pressure monitoring (Fig. 2).

Blood-pressure monitoring is a routine part of every surgical case, and it's critical in determining the surgical patient's hemodynamic status. Rapid blood-pressure changes, which can occur within a few heartbeats, may signal potentially serious complications. The proprietary tensymetry technology, on which the TL-150 is based, enables blood-pressure monitoring from the radial artery without using any external calibration source.

HOW IT WORKS
The medical staff externally applies the single-use, disposable pressure sensor after confirming the presence of a palpable pulse. Then the sensor is mated to the instrument's proprietary bracelet. Other than entering the patient's height and weight into the system, no further intervention is necessary.

Once the "zero paddle" is removed from the sensor, the system automatically compensates for atmospheric pressure and initiates a scanning operation to define an optimum pressure-sensing position over the radial artery. Next, the system performs a dynamic application pressure search. This involves slightly compressing the radial artery to determine a patient's mean arterial pressure, which is then scaled.

Tensys' design uses an actuator to move a sensor over the patient's wrist. The sensor has to float within a rigid frame attached to a serpentine arm designed to flex.

According to Stuart L. Gallant, chief technical officer for Tensys Medical, the TL-150 uses the patient's radial artery as the primary pressure source for the system. A proprietary algorithm determines diastolic and systolic pressures. A highfidelity beat-to-beat waveform is presented on either the TL-150's own display or an external display.

Proprietary servo-control algorithms enable the TL-150 to automatically and continuously respond to hemodynamic variations, as well as track significant blood-pressure changes without user intervention. The user is informed, via appropriate messaging and audible alerts, of any conditions that might affect patient safety, such as an abrupt loss of a signal.

"This device represents a significant improvement over the previously released TL-100. It is more userfriendly, requires fewer steps to gain an accurate waveform, and continues monitoring in more adverse conditions compared to its predecessor," says Dr. Gregory Janelle, assistant professor of anesthesiology at the University of Florida in Gainesville.

Anesthesiologists from the Emory University School of Medicine and the medical schools at the universities of Texas and Colorado also tested the TL-150 blood-pressure monitor and reported favorable results.

"With the TL-150, customers now have an accurate and non-invasive system that provides the superior control of beat-to-beat blood-pressure monitoring for cases where they currently use a traditional cuff," says Mike Martin, Tensys' chairman and CEO.

The TL-150 measures systolic blood pressure over a range of 30 to 245 mm Hg and diastolic blood pressure from 20 to 220 mm Hg. Its pulse-rate measurement capability is specified over a range of 30 to 250 beats per minute.




<-- prev. page     [1] 2     next page -->

Reprints     Printer-Friendly    Email this Article    RSS        Font Size     What's This?


  • Network-On-Chip Tools Arrive for The Masses
  • Tackling System Design Challenges Through Early Verification
  • ESL Tools Take Center Stage As Designers Move Up
  • Parasitic Extraction Tool Targets Next-Generation Custom ICs
  • Synopsys Jumps Into ESL-Synthesis Pool
  • Verify Control Systems Before Committing To Hardware
  • You're Using How Many FPGAs?
  • Tool Up For The FPGA Blitz
    1) Build A Smart Battery Charger Using A Single-Transistor Circuit
    (187 views today)
    2) Hot Hands For Some Cool Rock: Motion Sensing Meets Audio Engineering
    (170 views today)
    3) What's All This Transimpedance Amplifier Stuff, Anyhow? (Part 1)
    (91 views today)
    4) GPS-Derived Grandmaster Clock Delivers Ultra-Precise Time And Frequency Sync
    (90 views today)
    5) Downconverting Mixers Lower Power Consumption While Improving Performance
    (71 views today)
    ALL TOP 20







    POST YOUR COMMENTS HERE

    Name:

    Email:
    Rate this article:

     less useful more useful 
    1
    2
    3
    4
    5
    Your Comments:

    Enter the text from the image below




    Please refresh the page if you have trouble reading this text.
    (Acceptable Use Policy)
     
     

    PartFinder

    Find real-time pricing, stock status, same-day/next-day shipping options and more. Brought to you by Digi-Key. Go to PartFinder.    
    GlobalSpec

    PART SEARCH :
    Powered by: GlobalSpec - The Engineering Search Engine
    Sponsored Links

    Electronic Design Europe Electronic Design China EEPN Power Electronics Auto Electronics Microwaves & RF
    Mobile Dev & Design Schematics Find Power Products Military Electronics EE Events Related Resources