[Ideas For Design]
Precise Current-Transmitter/Source-Sink Calibrator Has Wide Compliance Range
Alfredo Saab,
Shasta Thomas
ED Online ID #20804
March 26, 2009
Copyright © 2006 Penton Media, Inc., All rights reserved. Printing of this document is for personal use only.
Reprints
Many industrial and control applications use analog
current loops as the physical link for measurements and actuator
control. Although several ranges are used, the most popular is still
the 4- to 20-mA standard.
Loop accuracy depends on the transmitter’s ability to transform
the sensor signal to a current. It also depends on the stability of the
current-transfer function (measured at the receiver) with respect
to temperature, physical length of the loop, stability of the loopsupply
voltage, and other uncertainties that appear as the hardware
deteriorates with age.
Besides linearity and stability of its current-transfer function,
the transmitter output must provide a large dynamic impedance
to minimize the influence of loop voltage (at the transmitter terminals)
on output current.
The maximum voltage tolerated at the terminals is also
important. The higher the betterbecause the electromagnetic
environment in industrial plants isn’t always benign. At the same
time, the minimum-allowed transmitter-terminal voltage, which
determines the maximum loop length for a given wire gauge (or
the minimum wire cross-section needed for a given loop run),
should be minimized.
The transmitter circuit in the figure meets these criteria. It
features a very high output impedance (108 Ω to 109 ) and a
wide compliance range (4 V to 90 V). The lower limit is set by the
amplifier’s minimum operating voltage, and the upper limit by
the power dissipation allowed in the output device (an n-channel
depletion MOSFET) at the high end of the current range (20
mA). For brief periods of 20 ms or so, the maximum voltage can go
as high as 200 V.
The basic circuit is configured as a loop calibrator (see the figure,
a). With the components shown, its accuracy from 0°C to 85°C is
10 bits, plus the added uncertainty of resistor R. Having only two
terminals, it can operate as either a sinking or a sourcing calibrator.
The loop calibrator can be modified for use as an interface to
an open-loop current sensor with a 0- to 4-V output range (see
the figure, b). The accuracy of the transfer function in this case
depends on the reference (MAX6138) and the precision of the
resistors in the interface network. Because the reference equipotential
(ground) is shared by the instrument and the transmitter,
you should exercise care not to create ground loops in the circuit
implementation.
The MAX4236 op amp’s input characteristics (20-V maximum
offset and 2-V/°C offset temperature drift), its high gain (110
dB), and its ability to operate with inputs down to the negative
rail make the influence of all other components negligible. To
achieve higher accuracy, therefore, you must use standard trimming
techniques that employ either mechanical or IC adjustment
potentiometers.
In particular, the amp’s ability to operate with input voltages
down to the negative rail is essential for circuit accuracy. That’s
because it allows the amplifier’s operating current to be forced
through the output current-sense resistor, where it becomes a part
of the output current under circuit control.
|